$0.00

  • *Wavelength Range

    • $
    • $

    Add-on Software

    Detailed Product Description:

    UV-VIS Spectrophotometer, 200-850 nm, direct attached cuvette light source

    Detailed Product Description:

    VIS-NIR Spectrophotometer, 350-1000 nm, direct attached cuvette light source

Key FeatureBest ForExample Applications
User Interchangeable SlitsUsers who wish to vary resolution and throughput during measurements, or switch from absorbance to fluorescence in minutesLife science and other labs using a variety of biological samples
Indicator LEDsConvenient visual reference for spectrometer operation and statusTeaching and general lab use
Thermal StabilityApplications that require repeatable results in industrial and other environments with varying temperatures and conditionsLED binning and light metrology, process monitoring
Reduced Unit to Unit VariationOEM applications such as uncalibrated sensitivity and other measurement needs where users benefit from low varianceOEM manufacturers of analytical instrumentation
ConfigurabilityOptimizing your spectrometer for application-specific requirements; adjusting range, throughput and resolution and adding features such as filters as requiredLaser characterization, low signal fluorescence and optimizing for specific absorbance bands
Plug & Play OperationUsers that want the convenience of simple, fast operation via the microUSB connection; enables users to take the measurement to the sampleRemote sensing measurements in the field, including air and water quality monitoring and solar irradiance
Ease of IntegrationOEMs and developers who need to integrate a spectrometer as part of a system via microUSB or RS-232Engineering labs, developers, OEM manufacturers; works with LabVIEW and other design platforms

UV-VIS

Detector:
Detector:
linear silicon CCD array
Pixels:
2048
Detector Collection Lens:
No
Electronics:
A/D Resolution:
16 bit
Connectors:
USB, high density 40 pin JAE DD4 connector
Inputs/Outputs:
8 x user programmable GPIOs
Trigger Modes:
4 modes
Illumination Source:
Tungsten halogen with a violet LED (~390-1000 nm)
Power Consumption:
250 mA @ 5 VDC
Spectroscopy:
Integration Time:
1ms – 65s
Optical Resolution:
1.33 nm FWHM (typical)
Wave Length Range:
350nm – 1μm
Dark Noise:
50 RMS counts
Dynamic Range:
1300:1 for a single acquisition
Input Fiber Connector:
SMA 905
Signal to Noise Ratio:
250:1 (full signal)
Stray light:
<0.10% at 435 nm

VIS-NIR

Specifications

Detector:

Detector:
linear silicon CCD array
Pixels:
2048
Detector Collection Lens:
No

Electronics:

A/D Resolution:
16 bit
Connectors:
USB, high density 40 pin JAE DD4 connector
Inputs/Outputs:
8 x user programmable GPIOs
Trigger Modes:
4 modes
Illumination Source:
Tungsten halogen with a violet LED (~390-1000 nm)
Power Consumption:
250 mA @ 5 VDC

Spectroscopy:

Integration Time:
1ms – 65s
Optical Resolution:
1.33 nm FWHM (typical)
Wave Length Range:
350nm – 1μm
Dark Noise:
50 RMS counts
Dynamic Range:
1300:1 for a single acquisition
Input Fiber Connector:
SMA 905
Signal to Noise Ratio:
250:1 (full signal)
Stray light:
<0.10% at 435 nm

Reviews

There are no reviews yet.

Be the first to review “Flame-CHEM”